- Ensure oiler function is enabled and Oil Pump Time in System Settings is sufficient.
- Ensure wand is fully submerged in oil. Replace container with full container if empty. Run oil prime function when new container used
- Check compressor pressure, it should be set to 90 PSI. If pressure is low, check for alarms that may be preventing the compressor from running.
- Check air pressure at system regulator, it should be between 60 -70 PSI.
- If above don’t fix, pump diaphragm is malfunctioning. Replace pump.